Decree 169: New Regulations on Medical Devices in Vietnam

On December 31, 2018, the Vietnamese government promulgated Decree No. 169/2018/ND-CP (Decree 169), amending Decree No. 36/2017/ND-CP (Decree 36) on medical device management. Decree 169 took immediate effect upon its issuance. The new regulations will have a great impact on the circulation of medical devices, especially foreign products, in Vietnam. Some of the major changes are set out below.

Classification of Medical Devices

Medical devices in Vietnam are classified and managed based on their potential risks, as determined by qualified entities. Under the previous regulation (Decree 36), risk classification results from qualified foreign entities could be recognized in Vietnam without conducting a separate local classification. This was beneficial to foreign manufacturers who sell their products in many countries, as it required no further time and expense for the local classification procedure in Vietnam and ensured a single classification result across multiple markets.

However, from January 1, 2019, foreign classification results are no longer accepted in Vietnam. Instead, the results must come from qualified local entities. It is worth noting that some local entities, despite being considered qualified by the Ministry of Health (MOH), still do not have appropriate capacity to conduct medical device classifications and, in practice, have issued classification results that were not accurate.

The classification rules under Decree 169 are generally stricter than under Decree36. For example, the final classification results will use the highest risk classification, regardless of whether the product has many use purposes or is combined with other medical devices with different risk classifications.

If an incorrect local classification result is submitted to the authority for product registration purposes, the authority will refuse to grant a product registration number if the number is not yet issued or revoke a registration number that has been granted, and may recall any circulating products.

Transitional Regulations

Decree 169 sets out the transition mechanisms as below.

  • Foreign classification results submitted to the authorities prior to December 31, 2018, will continue to be used for the registration procedure.
  • Local classification results issued before December 31, 2018, will be published on the MOH’s online system (http://dmec.moh.gov.vn) before April 1, 2019. MA license/import license holders are then required to review the relevant classification results in light of the classification rules under Decree 169 and report any incorrect classification to the MOH prior to July 1, 2019.

Registration of Medical Devices

Class A Medical Devices

A major change for Class A (lowest risk) medical devices is that the certificate of free sale (CFS) – not previously required – must now be included in any registration dossier submitted after December 31, 2018. This change may cause some difficulties for foreign manufacturers in terms of cost and time, and in some cases it may be impossible to obtain the CFS because some products that are considered Class A medical devices in Vietnam are not considered medical devices in other countries.

Previously, the granted registration of a Class A medical device could not be amended. If any change arose, a new registration procedure was required. However, Decree 169 stipulates a more flexible mechanism, in which the following changes can be made to the registration during the circulation of Class A medical devices.

  • Address of product owner or license holder.
  • Name of license holder.
  • Either name or address of product manufacturer.
  • Product packaging for in vitro medical devices.
  • Warranty establishments.
  • Labels and instructions for use, but the indication of the product cannot be changed.

Class B, C and D Medical Devices

A positive development for medical device companies is that Decree 169 allows a newly added form of registration, namely, “quick registration.” This will reduce the statutory time required to obtain a registration number from 60 days to 30 days, but is only applicable to the following situations:

  • The product has been circulated in at least two of the following countries: Japan, Canada, Australia, United States, or EU member country.
  • The product was circulated in Vietnam prior to December 31, 2018, and satisfies both of the following conditions: (i) The product was circulated for at least three years within the period of five years before the submission date, and (ii) there have been no warnings about the quality or safety of the product.

Additionally, from July 1, 2020, it will be required to include the ASEAN Common Submission Dossier Template (CSDT) in the submitted registration dossier for Class B, C and D medical devices. At that time, it will no longer be necessary to submit the technical document, instructions for use, and labels. It is expected that the MOH will issue further guidance on CSDT soon.

Decree 169 helps to simplify the customs clearance procedure by emphasizing that medical devices will be allowed to be imported into Vietnam without a written confirmation from the MOH that defines the imported product as a medical device when conducting customs clearance.

Transitional Regulations

In the transition time, it is worth noting the following regulations:

  • Class A medical devices are only allowed to be imported if they are registered.
  • A marketing authorization (MA) license for Class B, C, and D medical devices will only be required from January 1, 2020, onwards, instead of January 1, 2018, as set forth under Decree 36.
  • It is required to have a valid import license to import Class B, C, and D medical devices which are specifically listed out in Circular 30/2015/TT-BYT on Import of Medical Devices (Circular 30). Import licenses issued in 2018 (expiring on December 31, 2018) or 2019 will be valid until December 31, 2019.
  • Class B, C and D medical devices that are not specifically listed out in Circular 30 can continue to be freely imported into Vietnam without any import license, provided that their certificates of classification results are published on the MOH website http://dmec.moh.gov.vn.
  • For IVD biologicals (e.g., microbiology reagents, histology reagents, etc.), Decree 169 sets out specific regulations under which such IVD biologicals can still be treated as drugs under the Pharma Law.
  • For chemicals, insecticides and germicides for household and medical use, with only one indication of disinfecting medical devices, which have been granted MA licenses which expire after July 1, 2016 and before January 1, 2019, their MA licenses will be used until December 31, 2019.

RELATED INSIGHTS​ 

December 4, 2024
On October 28, 2024, Indonesia officially amended its existing Patent Law when the president ratified Law Number 65 of 2024. This comprehensive update—the third such amendment in the history of Indonesia’s Patent Law—introduces several key changes that will significantly impact patent protection and application processes in Indonesia. Key highlights and changes are outlined below. Definition of Invention The new law broadens the definition of “invention” to explicitly include systems, methods, and uses. Additionally, the law introduces formal definitions for traditional knowledge and genetic resources. Patentability Criteria Notable changes include: Computer programs are now excluded, with an exception for computer-implemented inventions. Theories and methods in science and mathematics are added to the list of excluded inventions. Previous restrictions on new uses of existing products are removed. Grace Periods The grace periods for some patent-related actions have been adjusted: The grace period for disclosures has been extended to 12 months (from 6 months previously), providing inventors with more flexibility in filing patent applications after initial disclosure. A newly introduced item is the grace period for a conventional patent application claiming priority rights, which is 4 months after the 12-month filing deadline under the Paris Convention. The grace period for annuity payments is 6 months (from 12 months previously) with a fine for late payments of 100% of the annual fee payable. Patent Holder Rights and Obligations Patent holders can now grant permissions to enforce patents. There is a new requirement for patent holders to submit annual statements on patent implementation in Indonesia. Compulsory Licensing Significant changes to compulsory licensing include: Establishment of licenses based on the principle of expediency. Limitations on license scope and transferability. Prioritization of domestic market needs. New provisions for technical improvements and economic significance. Government Patent Exploitation The new law contains specific provisions for the government’s implementation
December 4, 2024
Thailand Legal Basics, a valuable primer for foreign investors, explores all aspects of living and doing business in Thailand. Written by specialists at Tilleke & Gibbins in Bangkok, it is the only comprehensive English-language guide to the Thai legal system with a focus on the concerns of foreign business and investment.
December 2, 2024
Kratom (Mitragyna speciosa) was once listed as a prohibited substance under the Narcotics Act. Kratom was then removed from the narcotics list in 2021 and is now regulated under the Kratom Plant Act B.E. 2565 (2022), which allows the trading, use, import, and export of kratom. Furthermore, the Ministry of Public Health has allowed the use of kratom in food and drink, food supplements, and herbal products. Anyone who wishes to launch kratom food or herbal products must first obtain a product license from the Thai Food and Drug Administration (FDA). On September 4, 2024, the Thai FDA announced guidelines specifically for powdered kratom leaves and kratom leaf extracts (using water or ethanol as solvents) as ingredients in food supplements. These guidelines aim to provide business operators with a clear framework for launching kratom-based food products in Thailand. Key points from these guidelines include: Raw material specifications and standards: Powdered kratom leaves and kratom leaf extracts must contain at least the prescribed amount of mitragynine (one of the primary active compounds in kratom) as a marker of quality. For example, mitragynine must account for at least 1% by mass for powdered kratom leaves. Food supplement specification and standards: Food supplements containing powdered kratom leaves or kratom leaf extracts must not have a dosage of over 0.2 mg/day of mitragynine (including 7-hydroxymitraglynine, which may occur naturally), or 1 mg/day for food supplements containing kratom extracts that use water as a solvent. Addition of 7-hydroxymitraglynine to food supplements is prohibited. Mitragynine content in food supplement products must not exceed the specified dosage limit. Additionally, the food supplement product must be in a ready-to-eat form (e.g., liquid, gel, tablet, or capsule). If powdered kratom leaves or kratom leaf extracts are to be used as ingredients in food categories other than food supplements,
October 1, 2024
The Indonesian food and drug authority, also known as “BPOM,” issued a draft regulation on September 9, 2024, proposing standard labeling disclosing the sugar, salt, and fat content of packaged food products sold in the country. The draft Regulation on Nutritional Value Information on Food Labels to implement Government Regulation No. 28 of 2024 on the Implementation of the Health Law seeks to mandate “Nutri-Level” front-of-pack nutrition labeling to indicate the amount of sugar, salt, and fat at four possible levels, with a modified stoplight-color system: Level “A” (lowest amount) has a dark green background Level “B” has a light green background Level “C” has a yellow background Level “D” (highest amount) has a red background These levels are shown in the following sample image: The requirements for sugar, salt, and fat content for each level are based on amounts per 100 milliliters of ready-to-eat processed food as follows: Further requirements relating to nutrients reflected in the Nutri-Level labeling include: Foods labeled as level A are not allowed to contain natural or artificial sweetening food additives, either through direct addition or carried over from other ingredients. Foods labeled as level B may only contain natural sweetening food additives. Foods labeled as level C or D may use natural and artificial sweetening food additives. “Sugar” includes all monosaccharides and disaccharides, excluding lactose. Processed plain liquid milk and plain milk powder are exempt from the requirement to declare total fat content on the Nutri-Level label. Labeling Implementation The implementation of Nutri-Level labeling must comply with the following requirements: The Nutri-Level label must list all four letters (as shown in the examples below) unless the package meets the criteria for display of a simplified format label. The Nutri-Level of the processed food must be indicated by enlarging the relevant letter, as shown