Decree 169: New Regulations on Medical Devices in Vietnam

On December 31, 2018, the Vietnamese government promulgated Decree No. 169/2018/ND-CP (Decree 169), amending Decree No. 36/2017/ND-CP (Decree 36) on medical device management. Decree 169 took immediate effect upon its issuance. The new regulations will have a great impact on the circulation of medical devices, especially foreign products, in Vietnam. Some of the major changes are set out below.

Classification of Medical Devices

Medical devices in Vietnam are classified and managed based on their potential risks, as determined by qualified entities. Under the previous regulation (Decree 36), risk classification results from qualified foreign entities could be recognized in Vietnam without conducting a separate local classification. This was beneficial to foreign manufacturers who sell their products in many countries, as it required no further time and expense for the local classification procedure in Vietnam and ensured a single classification result across multiple markets.

However, from January 1, 2019, foreign classification results are no longer accepted in Vietnam. Instead, the results must come from qualified local entities. It is worth noting that some local entities, despite being considered qualified by the Ministry of Health (MOH), still do not have appropriate capacity to conduct medical device classifications and, in practice, have issued classification results that were not accurate.

The classification rules under Decree 169 are generally stricter than under Decree36. For example, the final classification results will use the highest risk classification, regardless of whether the product has many use purposes or is combined with other medical devices with different risk classifications.

If an incorrect local classification result is submitted to the authority for product registration purposes, the authority will refuse to grant a product registration number if the number is not yet issued or revoke a registration number that has been granted, and may recall any circulating products.

Transitional Regulations

Decree 169 sets out the transition mechanisms as below.

  • Foreign classification results submitted to the authorities prior to December 31, 2018, will continue to be used for the registration procedure.
  • Local classification results issued before December 31, 2018, will be published on the MOH’s online system (http://dmec.moh.gov.vn) before April 1, 2019. MA license/import license holders are then required to review the relevant classification results in light of the classification rules under Decree 169 and report any incorrect classification to the MOH prior to July 1, 2019.

Registration of Medical Devices

Class A Medical Devices

A major change for Class A (lowest risk) medical devices is that the certificate of free sale (CFS) – not previously required – must now be included in any registration dossier submitted after December 31, 2018. This change may cause some difficulties for foreign manufacturers in terms of cost and time, and in some cases it may be impossible to obtain the CFS because some products that are considered Class A medical devices in Vietnam are not considered medical devices in other countries.

Previously, the granted registration of a Class A medical device could not be amended. If any change arose, a new registration procedure was required. However, Decree 169 stipulates a more flexible mechanism, in which the following changes can be made to the registration during the circulation of Class A medical devices.

  • Address of product owner or license holder.
  • Name of license holder.
  • Either name or address of product manufacturer.
  • Product packaging for in vitro medical devices.
  • Warranty establishments.
  • Labels and instructions for use, but the indication of the product cannot be changed.

Class B, C and D Medical Devices

A positive development for medical device companies is that Decree 169 allows a newly added form of registration, namely, “quick registration.” This will reduce the statutory time required to obtain a registration number from 60 days to 30 days, but is only applicable to the following situations:

  • The product has been circulated in at least two of the following countries: Japan, Canada, Australia, United States, or EU member country.
  • The product was circulated in Vietnam prior to December 31, 2018, and satisfies both of the following conditions: (i) The product was circulated for at least three years within the period of five years before the submission date, and (ii) there have been no warnings about the quality or safety of the product.

Additionally, from July 1, 2020, it will be required to include the ASEAN Common Submission Dossier Template (CSDT) in the submitted registration dossier for Class B, C and D medical devices. At that time, it will no longer be necessary to submit the technical document, instructions for use, and labels. It is expected that the MOH will issue further guidance on CSDT soon.

Decree 169 helps to simplify the customs clearance procedure by emphasizing that medical devices will be allowed to be imported into Vietnam without a written confirmation from the MOH that defines the imported product as a medical device when conducting customs clearance.

Transitional Regulations

In the transition time, it is worth noting the following regulations:

  • Class A medical devices are only allowed to be imported if they are registered.
  • A marketing authorization (MA) license for Class B, C, and D medical devices will only be required from January 1, 2020, onwards, instead of January 1, 2018, as set forth under Decree 36.
  • It is required to have a valid import license to import Class B, C, and D medical devices which are specifically listed out in Circular 30/2015/TT-BYT on Import of Medical Devices (Circular 30). Import licenses issued in 2018 (expiring on December 31, 2018) or 2019 will be valid until December 31, 2019.
  • Class B, C and D medical devices that are not specifically listed out in Circular 30 can continue to be freely imported into Vietnam without any import license, provided that their certificates of classification results are published on the MOH website http://dmec.moh.gov.vn.
  • For IVD biologicals (e.g., microbiology reagents, histology reagents, etc.), Decree 169 sets out specific regulations under which such IVD biologicals can still be treated as drugs under the Pharma Law.
  • For chemicals, insecticides and germicides for household and medical use, with only one indication of disinfecting medical devices, which have been granted MA licenses which expire after July 1, 2016 and before January 1, 2019, their MA licenses will be used until December 31, 2019.

RELATED INSIGHTS​ 

June 17, 2026
Thailand’s new labeling requirements for medical devices, which include for the first time a unique device identification (UDI) requirement for software as a medical device (SaMD), take effect on June 20, 2026. The Notification of the Ministry of Public Health regarding Criteria, Methods, and Conditions on Labeling and Instructions for Use for Medical Devices 2025, which replaces a similar notification from 2020, was published in the Government Gazette on December 22, 2025. To ensure clarity, modernity, and patient safety, the regulation requires domestic manufacturers and importers to provide labels and instructions for use (IFU) that are clearly legible, complete, and free of false or misleading claims. It also permits IFU to be provided in electronic format, such as via QR codes, websites, or other digital channels—directly relevant to SaMD, where physical labels are impractical and electronic presentation is the natural medium. The notification distinguishes two categories for labeling language. Home-use medical devices (for lay users outside healthcare facilities) must have labels and IFU in Thai. Professional-use medical devices may display labels and documentation in either Thai or English. This distinction is significant for SaMD developers: software intended for clinical professionals may use English-language interfaces and IFU, while consumer-facing health applications must provide Thai-language content. Labeling and UDI Requirements Labels and IFU must include, at a minimum: Product name and intended purpose Quantity or volume Name and address of domestic manufacturer or importer Thai FDA approval number Lot, version, or serial number Manufacturing date and expiry date For SaMD, the version number requirement is particularly relevant. The regulation also mandates display of a UDI code for SaMD in risk category 2 (moderate-risk), category 3 (moderate- to high-risk), and category 4 (high-risk), according to Thailand’s medical device risk classification system (which complies with the ASEAN Medical Device Directive and the EU
June 5, 2026
On May 25, 2026, Vietnam’s Ministry of Health issued Circular No. 16/2026/TT-BYT governing free-of-charge medicine support programs for medical establishments (Circular 16). Circular 16 will take effect on July 10, 2026, replacing Circular No. 31/2018/TT-BYT, which currently regulates the same subject matter. Circular 16 introduces several significant changes compared to the existing legal framework. Removal of Prior Approval Requirement Under the current regulations, free-of-charge medicine support programs are divided into two categories: (1) entirely free-of-charge provision of medicines for all types of drugs and (2) partially free-of-charge provision applicable only to brand-name drugs under patent protection or drugs whose generic products with identical active ingredients and dosage forms are available in Vietnam. Under the current regulations, partially free-of-charge programs are subject to mandatory registration with the competent authority, while entirely free-of-charge programs could be implemented without prior approval. A key reform under Circular 16 is that it stipulates only entirely free-of-charge medicine support programs applicable to all types of medicines, thereby eliminating the partially free-of-charge category. In addition, free-of-charge medicine support programs may be carried out solely based on a written agreement between the pharmaceutical company and the medical establishment, without any requirement for prior approval from competent authorities prior to implementation. Written Agreement Requirements Circular 16 requires the pharmaceutical company and medical establishment to enter into a written agreement in accordance with a prescribed template. This agreement must include the following compulsory information: Information on the supported medicines Form of support (entirely free-of-charge provision to patients) Quantity of medicines provided Target patient groups and applicable indications Duration of the program Rights and obligations of each party Transitional provisions on the protection of patients’ rights upon completion of the program The agreement may contain other contents as agreed by the parties, provided that these do not contradict applicable laws.
June 4, 2026
Indonesia’s Minister of Health has issued Decree No. HK.01.07/MENKES/301/2026 on the Affixation of Nutritional Labels and Health Information to Ready-to-Eat Processed Food Products. The decree came into force on April 14, 2026, and was issued to implement the Health Law and Minister of Health Regulation No. 3 of 2026 on Disease Control. The decree requires the inclusion of Nutri-Level labeling on the front-of-pack nutrition labeling (FOPNL) to indicate the product’s nutritional level based on the content of sugar, salt, and fat (“gula, garam dan lemak (GGL)”). Changes from 2024 Draft Regulation The Nutri-Level labeling was previously proposed in 2024 by the Indonesian FDA (BPOM) through a draft regulation concerning nutrition information. While the categories of Nutri-Level labeling remain the same in the issued decree, the content requirements of sugar, salt and fat in the decree are different from the earlier proposal introduced in the 2024 draft BPOM regulation. In addition, the decree has further specified that the content of fat in the Nutri-Level labeling is the content of saturated fat, not total fat as previously proposed in the 2024 draft. The decree requires Nutri-Level labeling to be implemented in beverage products, which is the same as previously proposed in the 2024 draft BPOM regulation. Other food products may gradually become subject to mandatory Nutri-Level labeling under future implementing regulations. Nutri-Level Labeling Food levels as shown by the Nutri-Level labeling are classified into four color-coded categories from A to D: Level A (lowest amount) in dark green Level B in light green Level C in yellow Level D (highest amount) in red The Nutri-Level labeling is represented by the following image. The requirements for each level for sugar, salt, and fat content, based on amounts per 100 milliliters of product in beverage form, are as follows. Nutri-Level information must be
May 29, 2026
Indonesia’s Food and Drug Authority (BPOM) has issued Regulation No. 7 of 2026 on Drug Promotion and Advertising, establishing an updated framework for promotional activities involving medicinal products in Indonesia. The regulation took effect on April 16, 2026, and supersedes BPOM Regulation No. 2 of 2021 on Drug Advertising Supervision. The new regulation maintains general principles for advertising content, including requirements that advertisements be objective, complete, and not misleading, as further detailed in its annex. It also confirms that advertisements for nonprescription drugs directed to the public must obtain prior approval from BPOM before publication and must be in Bahasa Indonesia. The regulation provides a more comprehensive framework governing how drug promotion is conducted, introducing several notable additions and changes, as described below. Procedure and Requirements for Drug Advertisement Approval To apply for a drug advertisement approval, applicants must create an online account through the dedicated portal SIAPIK. Advertisement approval is available only for registered drugs; unregistered drugs are not eligible for advertisement approval with BPOM. The application must include the advertisement design, along with a translation if the design contains any wording in a foreign language. The submission format varies by media type, requiring, for example, copies in the form of print advertisements for visual media, scripts for audio media, and storyboards for audiovisual media. For online media—including social media—submissions should include any captions, descriptions, and hashtags that form an integral part of the advertising material. The approval timeline takes approximately 3–4 months, as BPOM will generally request additional information or revisions during the verification and evaluation process. Applicants have 20 days to submit any such requested documents. BPOM may also conduct a reevaluation of advertisements that have already received approval, based on monitoring results or new information regarding the safety and quality of the advertised drug. This